Troubleshooting - BRASSELER USA BSP MAX II Instructions For Use Manual

Sagittal/oscillator saw
Hide thumbs Also See for BSP MAX II:
Table of Contents

Advertisement

Troubleshooting

Symptom
Handpiece does not
start.
Handpiece lacks
power.
Handpiece stops
during use.
Handpiece runs after
trigger is released.
Handpiece will not
accept or retain saw
blade.
12
Sagittal/Oscillator Saw – PM-X12-700
Potential Cause
Trigger is in safe mode.
Battery pack is not installed.
Battery pack is not installed fully.
Battery pack is discharged.
Trigger is activated too slowly.
Handpiece has electrical/mechanical
malfunction.
Battery pack is discharged.
Battery pack is malfunctioning.
Variable speed throttle is out of adjustment. Send handpiece to Brasseler U.S.A. Medical, LLC, for service.
Battery pack is discharged.
Circuit protection is activated due to
excessive loading.
Trigger is malfunctioning.
Incorrect blade used.
Solution
Rotate trigger to run mode.
Install battery pack.
Remove and reinstall battery pack.
Charge battery pack.
Increase activation rate.
Send handpiece to Brasseler U.S.A. Medical, LLC, for service.
Charge battery pack.
Replace battery pack.
Charge or replace battery.
Release trigger to reset circuit protection.
Send handpiece to Brasseler U.S.A. Medical, LLC, for service.
Use only Brasseler USA® saw blades.