Handpiece Operation

• Rotate trigger to the SAFE mode.
• Install battery pack and saw blade.
• Position the sagittal head at the desired cutting angle
by retracting the sagittal release collar and rotating the
head.
• If using cutting guide, place blade in guide prior to
activation and use irrigation.
• Rotate trigger to the RUN mode.
• Depress trigger to operate the handpiece.
• If power loss occurs during use, replace the battery
pack with a fully charged sterilized battery pack.
• Rotate trigger to the SAFE mode when not in
operation.
Warning:
• Irrigation is required during resection in order to
prevent heat buildup which could cause bone
necrosis.
Sagittal Release
)
RUN Mode
Trigger*
SAFE Mode
*Shown in RUN position
Collar