BRASSELER USA BSP MAX II Instructions For Use Manual

Four station modular battery charger, 4-bay power unit, 110/230v, standard battery pack charging bay
Hide thumbs Also See for BSP MAX II:

Advertisement

SURGICAL POWER & ACCESSORIES
Rx Only
PM-X00-520/PM-X00-522/PM-X00-731-IR-12.13

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BSP MAX II and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BRASSELER USA BSP MAX II

  • Page 1 SURGICAL POWER & ACCESSORIES Rx Only PM-X00-520/PM-X00-522/PM-X00-731-IR-12.13...
  • Page 2: Table Of Contents

    Table of Contents Introduction Page 4 Intended Use Page 4 Warnings Page 5 Cautions Page 6 Explanation of Symbols Page 6 Features Page 7 Charging Bay – PM-X00-731 Page 8 Installation and Removal Page 8 4-Bay Power Unit – PM-X00-520 & PM-X00-522 Page 9 4-Bay Power Unit and Operation Page 9...
  • Page 3: Introduction

    Introduction Thank you for choosing Brasseler USA® Surgical Power & Accessories as supplier of your large bone power system. The information and procedures described in this manual are intended to assist medical professionals in the safe and effective use, care, cleaning, sterilization and long-term maintenance of BSPMAX™ II Large Bone Power System.
  • Page 4: Warnings

    Portable and Mobile RF communications equipment Charging Bay. can affect the function of the 4-Bay Power Unit. • Always use the Brasseler USA® charging bay when • Under certain classifications of risk, the World charging battery packs. Failure to comply may Health Organization (WHO), or local regulatory result in patient and/or medical staff injury.
  • Page 5: Cautions

    Cautions • DO NOT sterilize the 4-Bay Power Unit or Charging • REMOVE battery packs from Charging Bays when Bays. 4-Bay Power Unit is off to avoid battery discharge and/or irreparable damage to battery cells. • DO NOT connect 4-Bay Power Units in series. Explanation of Symbols Catalog number.
  • Page 6: Features

    Features Indicator Lights CHARGE (Yellow) READY (Green) SERVICE (Red) Charging Bay Cover Charging Bay Power Connection Main power switch Off/On Power Cord Connection Fuse holder • Power Connection – Connects and disconnects Power Unit from facility power. Power connection contains: - Fuse Holder - Main Power Switch - Power Cord Connection...
  • Page 7: Charging Bay - Pm-X00-731

    Charging Bay – PM-X00-731 The 4-Bay Power Unit can be configured with a combination of Charging Bay(s) and Bay Cover(s). The 4-Bay Power Units (PM-X00-520 and PM-X00-522), Standard Battery Pack Charging Bay (PM-X00-731) and Charging Bay Cover (PM-X00-521) are sold separately. The power unit charges up to four battery packs simultaneously. Cautions: •...
  • Page 8: 4-Bay Power Unit - Pm-X00-520 & Pm-X00-522

    4-Bay Power Units – PM-X00-520 (110V) & PM-X00-522 (230V) The 4-Bay Power Unit can be configured with a combination of Charging Bay(s) and Bay Cover(s). The 4-Bay Power Units (PM-X00-520 and PM-X00-522), Standard Battery Pack Charging Bay (PM-X00-731) and Charging Bay Cover (PM-X00-521) are sold separately.
  • Page 9: Battery Pack Charging

    Battery Pack Charging 1. Insert battery pack into a charging bay receptacle. Battery Pack 2. Follow warnings, cautions and notes in BSPMAX™ II PM-X00-710 (9.6V) & PM-X00-715 (12V) Battery Charging Packs Instructions for Use. Receptacle 3. A single light will illuminate to indicate battery pack status.
  • Page 10: Troubleshooting

    Troubleshooting Symptom Potential Cause Solution Fans and indicator lights do not function. Power cord not connected. Connect power cord. Power unit is turned OFF (O). Turn power switch On (-). Fuses are open/blown or missing. Replace with appropriate rated fuses. See 4-Bay Power Unit fuse replacement.
  • Page 11 Troubleshooting (continued) Symptom Potential Cause Solution Charging bay will not fully Misalignment of electrical plug. Send power unit and charging bay to Brasseler U.S.A. connect onto power unit. Medical, LLC, for service. Electrical plug blades bent. SendchargingbaytoBrasseler U.S.A. Medical, LLC, forservice. Indicator lights remain illuminated Charging bay malfunctioned.
  • Page 12: Care And Maintenance

    Care & Maintenance Brasseler USA® Surgical Power & Accessories Follow a regular care regimen that includes routine recommends that all BSPMAX™ II components cleaning and a thorough inspection for damage. (handpieces, attachments and accessories excluding Routine preventive maintenance performed every battery packs) be returned to Brasseler U.S.A.
  • Page 13: Cleaning Recommendations

    Cleaning Recommendations General Cleaning Precaution: Cautions: • Follow universal precautions and protective apparel • DO NOT immerse the 4-Bay Power Unit or when handling and cleaning contaminated Charging Bays in liquid. instruments. • DO NOT use solvents, lubricants, or other chemicals to clean the 4-Bay Power Unit or Charging Warnings: Bays unless otherwise directed.
  • Page 14 Cleaning Recommendations (continued) Cleaning Procedures: 1. Set the main power switch to OFF(O). 6. Clean charging contacts using a cotton swab lightly 2. Remove all battery packs. with isopropyl alcohol. DO NOT clean the charging 3. Disconnect the 4-Bay Power Unit power cord from contacts with abrasives.
  • Page 15: Specifications

    Specifications 1. Performance PM-X00-520 PM-X00-522 Input Voltage 100VAC – 120VAC 220-240VAC Max. Rated Current Frequency 50 – 60 Hz 50 – 60 Hz Output Voltage Max. 15VDC 15VDC Classification Protection Class 1 Protection Class 1 Class of Rating Continuous operation Continuous operation Socket Outlet 5A 100 –...
  • Page 16 Specifications (continued) 3. Compliance Standards EMC Compliance Standards: DN60601-1-2; IEC60601-1; CSA 60601.1 Safety Compliance: EN60950-1; IEC60950-1; CSA 60950-1; UL 60950-1 4. Environmental Requirements Operating: Transport : • Ambient temperature : 50°F to 70°F (10°C to 21°C) • Ambient temperature : -4°F to 158°F (-20°C to 70°C) •...
  • Page 17 Specifications (continued) 5. Electromagnetic Compatibility Requirements Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The BSPMAX™ II battery charger is intended for use in the electromagnetic environment specified below. The user of the BSPMAX™ II battery charger should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance...
  • Page 18 Specifications (continued) 5. Electromagnetic Compatibility Requirements continued Guidance and Manufacturer’s Declaration – Electromagnetic Emissions Immunity test IEC 60601 Compliance Electromagnetic environment - guidance test level level Portable and mobile RF communications equipment should be used no closer to any part of the BSPMAX™ II Battery Charger, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
  • Page 19 Specifications (continued) 5. Electromagnetic Compatibility Requirements continued Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The BSPMAX™ II Battery Charger is intended for use in the electromagnetic environment specified below. The user of the BSPMAX™ II Battery Charger should assure that it is used in such an environment. Immunity test IEC 60601 Compliance...
  • Page 20 Specifications (continued) 5. Electromagnetic Compatibility Requirements continued Guidance and Manufacturer’s Declaration – Electromagnetic Emissions Immunity test IEC 60601 Compliance Electromagnetic environment - guidance test level level Voltage dips, short <5%U 100% Reduction interruptions and (>95% dip in U (10 ms) voltage variations on For 0.5 cycle power supply input...
  • Page 21 Specifications (continued) 5. Electromagnetic Compatibility Requirements continued Guidance and Manufacturer’s Declaration – Electromagnetic Emissions Recommended separation distances between portable and RF communications equipment and the BSPMAX™ II handpiece. The BSPMAX™ II Battery Charger is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
  • Page 22: Repair Service

    BSPMAX™ II Large Bone Power System. Brasseler USA® Surgical Power & Accessories warrants any service or repair work performed will be free from defects in material or workmanship for the period of ninety (90) days from date of service or repair. This warranty applies to the actual work performed.
  • Page 23: Warranty

    Brasseler USA® Surgical Power & Accessories.
  • Page 24: Return Goods Policy

    Note: It is unlawful to ship contaminated non-sterilized products. Contact a Customer Service Representative at Brasseler USA® Surgical Power & Accessories at 800-535-6638 to obtain a “Return Merchandise Authorization” (RMA) number. Please include the following information with the returned product(s): •...
  • Page 25: Product Disposal

    STERILIZATION CASE – 3 HANDPIECES PM-X08-000 1/4 IN JACOBS ® CHUCK KEY PM-X08-001 5/32 IN JACOBS ® CHUCK KEY BRASSELER USA® SURGICAL POWER & ACCESSORIES OFFERS A COMPLETE LINE OF CUTTING ACCESSORIES (SAW BLADES/RASPS, BURS, K-WIRES, STEINMANN PINS, TWIST DRILLS AND ORTHAPAEDIC PIN PACKS).
  • Page 26 Inc. Brasseler U.S.A. Medical, LLC, is not affiliated with any of the above and makes no claim to copyrights or trademarks which are the property of these companies. Brasseler USA® and BSPMAX™ are trademarks of Peter Brasseler Holdings, LLC, or its affiliates. All other trademarks are trademarks of their respective owners or holders.