Invacare® Storm⁴ Series
5. You see strip fuse (2) once fuse holder (1) is open.
6. If strip fuse has blown, you must first ascertain and
rectify the cause of the fault.
7. Main fuse may only be replaced once fault has been
rectified.
8. Remove strip fuse nuts (3).
9. Remove strip fuse.
Installing Main Fuse
1. Replace strip fuse.
2. Install parts in reverse order.
Make sure to press two parts of fuse holder
shroud together until it snaps.
If older fuse holder version is used, lock fuse
holder with UL94V0 tie wrap.
3. Test all functions.

6.10.6 Cable Routing

4 mm Allen key
Oblique pliers
Cable ties
Checking
1. Remove rear and centre shroud. See 6.6.1 Replacing
Rear Shroud, page 21 and 6.6.2 Replacing Centre Shroud,
page 22).
2. Pull batteries out of battery box. See Making Batteries
Accessible.
3. Check fuse cable (1) and battery cable (2) for visible
damage and crushing points.
4. Check all cables for visible damage and crushing
locations. Replace any damaged cables.
5. Pull each plug carefully. The plug should not come out
of socket.
6. If one of plugs is loose, press it back into socket with
light pressure. The plug must engage.
7. Make sure that plug is now firmly in socket. If not,
repeat previous operation.
8. Install parts in reverse order.
9. Test all functions.
48
Fig. 6-42
Fig. 6-43
1555830-C