2. Description

2.1 General

The Apollo Tec 3 Style vaporiser is designed for "out of circuit" use in fresh
gas supply of a continuous flow techniques of inhalation anaesthesia. Each
vaporiser is agent specific and is clearly labelled with the anaesthetic agent
that it is designed for.
The Apollo Tec 3 Style vaporiser is temperature, flow and pressure
compensated so that its output remains relatively constant despite cooling
due to the vaporisation and variations in inlet flow.
The Apollo Tec 3 Style vaporiser provides accurate concentrations of
anaesthesia gases in the fresh gas supply; the concentration is specified
using the dial on vaporiser. The fresh gas supply must be between 0.5 and
10LPM.
The Apollo Plug on Block Tec 3 Style vaporiser is designed to be used on a
Selectatec Compatibility Block only.
Warning: If the vaporiser has been inverted, connect to a gas
scavenging system, set the dial to the top output and purge
the system with the carrier gas at 5LPM for 5 minutes.
Warning: improper use may result in patient injury.
Caution: Ensure the vaporiser is kept upright at all times.
Caution: Turn the vaporiser to the OFF position when not in
use.
Doc. No. 157
20 March 2019
Issue 1.0 Draft