ensure that the filler or bottle is not damaged before use.
2.
Remove the filler cap by turning it counter-clockwise and insert the
bottle nozzle into the filler port. Make sure that the bottle filler
nozzle keys are aligned with the index slots in the filler port, as
illustrated in Fig. 4.
3.
Press the agent bottle fully into the port of the filler block. Fill the
vaporiser until the maximum level mark
level indicator. Pay attention to the air return bubbles flowing into
the agent bottle this is guidance that the liquid is flowing.
4.
Once the vaporiser is full and the continuous stream of bubbles has
stopped, release the agent bottle from the filler port.
5.
Remove the agent bottle and replace the filler cap and agent bottle
cap. Make sure the filler cap is tight to reduce any possible leaks.
Caution: If the vaporiser was dry before filling, the level will
decrease slightly as the wick absorbs the agent.
Figure 4 Filling a Quick Fill Filler Vaporiser
5.3.6 Quick Fill Filler (draining procedure)
Doc. No. 157
20 March 2019
Issue 1.0 Draft
is reached on the sight
Apollo Operations Manual
Page 20 of 27