Inspection for damage and wear.
Renewal of wicks, seals and any damaged, worn or outdated
components.
Lubrication where necessary.
Checks of delivered vapour concentration under closely defined
conditions at different
temperatures and flows with re-calibration or adjustment where
necessary.
Caution: Halothane vaporisers must be serviced annually.

7.2 Cleaning

Warning: Do NOT put water or any other solvent in the Apollo
Tec 3 Style vaporiser. Do NOT immerse the vaporiser in any liquid,
including water.
Warning: Do NOT autoclave the vaporiser.
Clean the exterior of the Apollo Tec 3 Style vaporiser with a cloth dampened
with CE marked disinfectant as per the manufacturer's instructions.
Never allow cleaning agents to accumulate in the filler or gas inlet and outlet
ports or around the dial control.
The Apollo Tec 3 Style vaporiser must be cleaned prior to each use.
Warning: If the vaporiser is filled with incorrect agent or any
other contaminant, drain vaporiser and return to service centre stating
the cause of contamination.
Doc. No. 157
20 March 2019
Issue 1.0 Draft
Apollo Operations Manual