Installation; General - Norvap Apollo Operation & Maintenance Manual

Table of Contents

Advertisement

4

Installation

The Apollo Tec 3 Style vaporiser must always be mounted between the flow
meter unit and the patient breathing circuit, but upstream of any absorber or
humidifier. The vaporiser must be connected so that the flow of gas to the
patient is as indicated by the arrows on the device.
Check the integrity of the fittings to ensure that they are leak tight. If in doubt,
seek advice from the manufacturer of the equipment to which the vaporiser is
attached.
Ensure that the vaporiser is cleaned prior to first use. Remove the vaporiser
from its packaging, clean the exterior of the Tec 3 style vaporiser with a cloth
dampened with CE marked disinfectant as per the manufacturer's
instructions.
Never allow cleaning agents to accumulate in the filler or gas inlet and outlet
ports or around the dial control.

4.1 General

Unless otherwise specified, all Apollo Tec 3 Style vaporisers are supplied as
standard with either 23mm Cagemount fitting inlet and outlet ports or a
Selectatec compatible ports.
Warning: Do not lift or support the vaporiser by the control dial, handle
with care at all times.
Warning: Do not use vaporiser if the liquid level is below the minimum
marker.
Warning: Allow the vaporiser to acclimatise to ambient temperature of
the location it is being used in, before the vaporiser is used.
Doc. No. 157
Issue 1.0 Draft
Apollo Operations Manual
20 March 2019
Page 10 of 27