Figure 2 Filling a screw cap filler vaporiser
4.
Check that the agent to be used is the same as that specified on
the front of the vaporiser. In a ventilated area, pour the anaesthetic
agent slowly into the filler opening, observing the agent level
through the agent level indicator.
Caution: If the vaporiser was dry before filling, the level will decrease
slightly as the wick absorbs the agent.
5.
When the agent level reaches the maximum level mark
agent level indicator the vaporiser is full. Replace the screw cap by
turning it clockwise. To prevent leakage, ensure that the screw cap
is fully tightened.
5.3.2 Screw Cap Filler (draining procedure)
Caution: Do not allow the drainage container to become completely full
during draining.
1.
Remove the screw cap to reveal the drain plug.
2.
Position a properly labelled container under the drain spout.