Caution: Make sure that the locking lever is pushed all the way down
before turning it. The mechanism can be damaged if the locking lever is
turned before it is at the full extent of its vertical travel.
Warning: To help to ensure correct operation, do not use a vaporiser
which is visibly out of line on the manifold or which can be lifted off the
manifold when the locking lever is in the locked position.
6. Push the locking lever all the way down and turn the lever a quarter turn
clockwise, to lock the vaporiser onto the manifold.

4.4 Vaporiser Removal from the Manifold

Turn the dial to the 'OFF' position, if the vaporiser is not completely
1.
turned 'OFF' then it cannot be released from the manifold.
2.
Turn the locking lever a quarter of a turn anti-clockwise to the
unlocked position which will release the vaporiser from the
manifold.
3.
Carefully lift the vaporiser up from the manifold, handling it from the
main body.

5 Operating Instructions

5.1 General

Warning: Do not fill the vaporiser until the dial is in the 'OFF'
position, do not turn the dial whilst filling or attempt to over fill
the vaporiser above the
Warning: Do not fill vaporiser with any other agent other than
the agent specified vaporiser labelling. The vaporiser is
designed for labelled specific agents only and filling of any
other agent can prove to be dangerous to the patient.
Doc. No. 157
20 March 2019
line.
Issue 1.0 Draft
Apollo Operations Manual
Page 12 of 27