6.6 32-Point Line arisation
P47: - - - a
Linearisation
Linearisation is the method of assigning requested (calibrated or calculated) level, volume or flow to values measured by the transmitter.
It can be used for instance if the sound velocity is not known (LEVEL⇒LEVEL) or in the case of tank with other shape than under 6.4 or open
channel other than under 6.5 (LEVEL ⇒ VOLUME or LEVEL ⇒ FLOW).
a
0
1
P48:
Linearisation table
Data-pairs of the linearisation table are handled in a 2x32 matrix, consisting of two columns.
Left column "L"
LEVEL measured
The left column values (indicated on the display as "L") contain the measured LEVEL values.
The right column values (indicated on the display as "r") contain the calibrated values and are interpreted according to the selected measurement
value in P01(a).
Number of data pairs
in the table
SE/SG-300 series two-wire compact ultrasonic level transmitters
LEVEL or VOLUME or FLOW to be transmitted and displayed
Address of data pair
Enter
the table
Exit the
+
table
Linearisation
OFF (FACTORY DEFAULT)
ON
Right column "r"
Level
Right
Level
measured
hand
volume
Left hand
column
value
column
flow
Entering values
+
Copying measured value
+
Neglect modification (CANCEL)
+
Entering "0" (closing up the tble)
+
Canceling last modification
}
to be transmitted
and displayed
39