Cannon TE-3000 Instruction & Operation Manual page 27

Thermo electric temperature bath
Table of Contents

Advertisement

NOTE
Selecting a report interval
If the user selects to enter a 1 from the COMMUNICATION SETUP
menu, the COMM PORT SPEED menu will display five baud rate
options: 1200, 2400, 9600, 19.2K, and 38.4K BAUD. The baud rate
selected will be effective on both the RS-485 interface (polled) and the
RS-232 interface (point-to-point). The user may enter a number from 1
through 5 followed by ENTER or else the MENU key may be depressed
to return to the normal display.
The number shown after the colon denotes the present baud rate.
b. Configuring the mode of operation
If the user selects to enter a 2 from the COMMUNICATION SETUP
menu, one of the two following submenu screens will be shown. If the
TE-3000 was previously configured to operate in a polled (RS-485)
environment, the first display will be provided - showing that timed
reports are disabled and the present mode is polled (:1). If the user wishes
to operate in a point-to-point environment with or without timed interval
reports, a 2 should be entered now.
If the TE-3000 was previously configured to operate in a point-to-point
(RS-232) environment, the following display will be provided - showing
that timed reports may be utilized and the present mode is point-to-point
(:2). An entry of 1 at this time will place the bath in a polled environ-
ment. For an entry of 3, please see the next section (5.D.) for the re-
sponse to a TR command.
c. Configuring reporting periods
If a 3 is entered in the COMMUNICATION MODE menu, the system
was previously configured for point-to-point operation and therefore the
following display is shown. The user may select any reporting interval in
"seconds" or NO reporting interval which is OFF. The :1 in the bottom
right hand corner indicates the presently selected period. A number entry
from 1 to 8 is acceptable.
CANNON
®
TE-3000 Thermoelectric Constant Temperature Bath
Version 1.1c — May, 2005; CANNON
2139 High Tech Road • State College, PA 16803 • USA
COMM PORT SPEED
1. 1200
4. 19.2K
2. 2400
5. 38.4K
3. 9600
COMMUNICATION MODE
1. Polled
2. Point to Point
(Reports Disabled):1
COMMUNICATION MODE
1. Polled
2. Point to Point
3. Config Reports :2
®
Instrument Company
:3
23