Digital flat-panel x-ray detector
Table of Contents

Advertisement

Table2
This Wireless Detector is intended for use in the electromagnetic environment specified below. The
customer or the user of this Wireless Detector should assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge
(ESD)
IEC61000-4-2
Electrical fast Transient
/ burst
IEC61000-4-4
Surge
IEC61000-4-5
Power frequency
(50/60Hz) Magnetic
field
IEC61000-4-8
Voltage dips, short
Interruptions and
Voltage variations on
power supply input lines
IEC61000-4-11
Note: Uт is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration- Electromagnetic immunity
IEC 60601 test level
±6 kV Contact
±6 kV Contact
±8 kV Air
±8 kV Air
±2kV for power supply
±2kV for power supply
lines
lines
±1kV for input/output lines
±1kV for input/output lines
±1 kV differential mode
±1 kV differential mode
±2 kV common mode
±2 kV common mode
3.0 A/m
3.0 A/m
<5% Uт (>95% dip in Uт)
<5% Uт (>95% dip in Uт)
for 0.5cycle
for 0.5cycle
40% Uт (60% dip in
40% Uт (60% dip in
Uт) for 5 cycle, 6
cycle
70% Uт (30% dip in
70% Uт (30% dip in
Uт) for 25 cycle, 30
cycle
<5% Uт (>95% dip in Uт
<5% Uт (>95% dip in Uт )
for 5 sec
) for 5 sec
46
Compliance level
environment – guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of this Wireless Detector
Uт) for 5 cycle, 6
requires continued
cycle
operation during power
mains interruptions, it is
Uт) for 25 cycle, 30
recommended that this
cycle
Wireless Detector be
powered from an
uninterruptible power
supply or a battery
Electromagnetic