Resetting The Volume Calibration - AND MPA-10 Instruction Manual

Single channel electronic pipette, multi channel electronic pipette
Table of Contents

Advertisement

11-2 Resetting the volume calibration

Go through the following procedure to restore the factory default settings for volume
calibration:
(μL calibration mode)
[ 3 ] Press the
mode.
[ 4 ] Press the
to the actual dispensed volume.
[ 5 ] Press the
dispensed volume.
[ 6 ] The notice mark
the volume calibration has been performed.
(In this example, the dispensed amount is
corrected and altered to 100μL.)
* After calibrating, a volume range that can be
selected may be limited depending on available
movement range of the piston.
(The display is example.)
[ 1 ] Press the
[ 2 ] Press and hold the
three seconds while "AUTO" is selected.
32
Enter key to enter μL calibration
Up or Down key to alter the value
Enter key to confirm actual
lights up to indicate that
Back key.
Back key for approx.