Method for mg calibration
1. Select the mg unit beforehand. (Refer to "11-3 Dispensing in a unit of weight (in
mg unit)")
2. Set the dispensing amount of the MPA / MPB series to the weight to which you
would like to calibrate it. (The example is 100mg)
3. Using an electronic balance, measure and record an actually dispensed weight as
opposed to the selected dispensing amount setting. (The example is 95mg)
4. Enter an actual dispensed weight to the pipette by the following procedure.
(mg calibration mode)
mg
[ 1 ] Press the
select an operation mode.
[ 2 ] Press and hold the
seconds while "AUTO" is selected.
[ 3 ] Press the
mode.
mg
In mg calibration mode, the Notice mark
illuminates while μL mark is turned off
[ 4 ] Press the
to the actual dispensed weight. (The example is
mg
95mg)
[ 5 ] Press the
dispensed weight.
(In this example, the dispensed amount is
mg
corrected and altered to 100mg.)
35
Back key to enter the mode to
Back key for approx. three
Enter key to enter mg calibration
Up or Down key to alter the value
Enter key to confirm the actual