Care of the fall detector
• Avoid getting the fall detector wet (if it does then
dry it quickly and make a manual test call to confirm
it is working correctly).
• If required, the fall detector may be cleaned
periodically with a damp cloth and mild detergent.
• It is recommended that you test the fall detector
once a month:
• Make a manual test call.
• Start with the fall detector vertical. Place the fall
detector horizontal and then tap within 3 seconds
after placing horizontal. After 5 Seconds the fall
detector should beep and the green light start to
flash. If you put the unit quickly back upright (in its
stand, for instance), the alarm call will not be raised.