Automatic fall detection
• Green light and two "beeps" indicate that the fall
detector has been activated.
• The fall detector will then monitor for between 6
and 15 seconds to verify if there could have been a
fall. The green light will flash while it does this.
• If the detector has decided that there is a fall, then a
red light and four "beeps" indicates that the fall
detector is about to trigger an alarm call (because it
has detected an impact and has not been upright
for between 6 and 15 seconds).
• If the detector has decided that it was activated by
error then the green flashing light will stop after
between 6 to 15 Seconds and the unit will revert to
normal again.
• Due to the complex dynamics of falling, all falls may
not be detected. In this unlikely event the manual
button can still be used to call for help.

Manual operation

• Press the button on the fall detector to raise a
manual alarm call.
• The fall detector will "beep" and a red light
indicates the fall detector is triggering an alarm
call.
• If the red light flashes on and off, the fall detector is
still triggering an alarm call, but is also indicating to
you that the battery needs to be replaced.
Remember, the fall detector will only detect you falling if
you are wearing it.