EMC (Electromagnetic Compatibility)
Receiver (recorder unit,
real time viewer, antenna
lead set)
Transmitter (capsule
endoscope)
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
•
In order to provide the intended functionality, the system
must emit RF energy while in operation. This may affect
electrical devices in the vicinity.
•
The system is equipped with a function for device
identification via radio transmission, and is thus subject to
electrostatic interference from other equipment in the vicinity.
Frequency range of possible
interference: 426.8 to 440.8MHz
Central frequency: 433.8MHz
Required bandwidth: 7MHz
Modulation method: Minimum shift
keying (MSK)
High frequency output: +37dBµV/m
(when inside body) phantom
Important Information - Please Read Before Use
Frequency range of possible
interference: 306 to 322MHz
(for US/Canada)
Central frequency: 315MHz
Required bandwidth: 7MHz
Modulation method: Minimum shift
keying (MSK)
High frequency output: +37dBµV/m
(when inside body) phantom
27