Battery Pack - Olympus EndoCapsule Instruction Manual

Capsule endoscope system
Table of Contents

Advertisement

Important Information - Please Read Before Use
About the recharger's power cord

Battery pack

24
For use in America and Canada
The following Power Cord should be used for this battery charger.
Type SPT-2, 2C/18AWG, UL Listed and CSA Certified cord set with
non-polarized plug, 125V, 10A
For use in Europe
The following Power Cord should be used for this battery charger.
Type H03VVH2-F~0.75mm
250V, 2.5A
For use in England
The following Power Cord should be used for this battery charger.
Type H03VVH2-F~0.75mm
250V, 3A
Do not disassemble or modify the battery pack. The battery
pack is equipped with a safety circuit and a protective circuit.
Disabling either or both of these circuits may result in
excessive heat, release of smoke, explosion, and/or a fire.
Do not connect the (+) and (-) terminals of the battery pack
using pieces of metal. In addition, do not store or transport
the battery pack alongside other metallic objects. The battery
pack may be short-circuited, causing an overflow of current,
resulting in excessive heat, release of smoke, explosion,
and/or a fire from the battery, or excessive heating of the
metallic item.
Do not heat the battery pack or bring it in contact with fire.
The insulating material may melt, gas could be released, the
safety circuit may be damaged, and electrolytic solution may
be ignited, causing excessive heat, release of smoke,
explosion, and/or a fire.
Do not use or store the battery pack near flames or a source
of high temperature (over 80°C/176°F) such as a stove. If the
plastic separator is damaged by heat, the battery pack will be
short-circuited internally, causing excessive heat, release of
smoke, explosion, and/or a fire.
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
2
, Europe Approved cord set with non-polarized plug,
2
, BSI certified cord set with non-polarized plug,