Directions For Use - Cardiac Science POWERHEART G3 PLUS 9390E AED Quick Reference Manual

Table of Contents

Advertisement

Getting Started

Directions for use

WARNING! Do not reuse pads.
Used pads may not adhere properly to the patient. Improper pad
adhesion may result in skin burns. Improper pad adhesion may result
in improper AED performance. Used pads may cause patient-to-
patient contamination.
WARNING! Reduced therapy delivery.
Failure to remove blue liner completely could impact therapy
delivery.
Caution. Short-term use only. Not for pacing.
DO NOT open defibrillation pads package until ready to use. Short
term use only.
Pads are not intended for use in pacing.
Caution. Equipment Damage.
Do not pull on the lead wire to separate the pads from the blue liner.
Note: Store pads at room temperature.
Note: Pads are intended for adult use.
1.
Ensure skin site is clean and dry.
2.
Tear open the foil package and remove pads.
3.
Separate one pad from the blue liner by peeling from the tabbed
corner.
Note: DO NOT pull on the lead wires.
4.
Place pad on skin in either location shown.
5.
Separate the second pad from blue liner by peeling from the
tabbed corner.
Note: DO NOT pull on the lead wires.
6.
Place the second pad on the opposite location as shown.
2-10
70-00915-01 F
Powerheart® AED G3 Plus 9390A and 9390E