When the measurement has stabilized, you can press the HOLD key to
freeze it and/or MEM to record it in memory. You can also use the
AUTO-HOLD function (see §2.4) to freeze the measurement
automatically when the clamp jaws are opened. This can be useful when
you have only one hand free to make a measurement. To facilitate
making measurements, you can configure an alarm on the current
measurement value (see §2.4).
2.4. CONFIGURATION (SET-UP)
Use the ▼ and ▲ keys to scroll through the screens of the SET-UP
menu.
Clamp-On Ground Tester Model 6418
Set the switch to SET-UP.
CLR (erase memory) menu.
Press the ► key to enter the CLR menu.
To cancel, press ►.
To erase all records, press and hold the ▲ and ▼ keys
simultaneously. The instrument emits 5 audible beeps before
erasing the memory.
SND menu (activation of the sound).
Press ► to enter the SND menu. Use ▲ and ▼ to display or
hide the
symbol.
When
is displayed, the instrument emits an audible
signal when keys are pressed and when alarm thresholds are
crossed.
Confirm by pressing ►.
During the measurements (when the switch is set to Ω+A or
A), a long press on ► activates/deactivates the audible
signal.
STOP menu (automatic standby).
Press ► to enter the STOP menu. Use ▲ and ▼ to display
or hide the
symbol (permanent operation).
When
is
displayed, automatic switching of the instrument to
standby at the end of 5 minutes is disabled. Confirm by
pressing ►.
ALΩ menu (impedance alarm).
Press ► to enter the ALΩ menu. Use ▲ and ▼ to scroll
through the display:
: no impedance alarm.
: impedance alarm activates when impedance >10Ω.
: impedance alarm activates when impedance <10Ω.
Press ►; the alarm threshold blinks. Set it using ▲ and ▼.
Acceptable values are between 1 and 199Ω. A long press
scrolls the values faster.
Confirm by pressing ►.
13