oticon REM AutoFit Quick Manual page 7

Hearing aid fittings

Advertisement

Before using REM AutoFit
Hearing aid settings during measurement
Step 4: Finish
The graph now displays the hearing aid gain measured in your client's ear,
and you can apply the adjustments from Step 3: Aided. You can choose
whether the manually fine-tuned gain should be included or not. The graphs
and the instruments update according to what you select, so you can dem-
onstrate each option to the client.
Select Apply REM AutoFit to apply automatic adjustments. REM AutoFit
data measured in previous fittings will be overwritten. Gain modifications
from the automatic adjustment are applied to all programs. If manual fine-
tuning is included, it is only applied to the current program and linked
programs.
7
Step-by-step Guide
Step 2: Unaided
Step 1: Calibrate
Automatic
Step 3: Aided
Manual
Select Discard All to discard REM AutoFit adjustments. The data in use before
you entered REM AutoFit is then restored.
Removing measured real ear data
If you no longer want to use the REM AutoFit adjustments and want to use
the predicted data instead, do as follows:
1.
Leave the REM tool if it is open.
Go to the Hearing Instrument menu on the menu bar and select
2.
Remove REM AutoFit Results. This option is only available if corrections
based on REM AutoFit are in use.
If you also want to remove Unaided measurement data, do as follows:
Open the REUG tool in the Selection step.
1.
2.
Change REUG to predicted or delete REUG data.
If you change the acoustics, REUG or RECD data after completing REM AutoFit,
the fitting software will represcribe using predicted real ear data (i.e. REM
AutoFit adjustments are removed). To use measured data, you must repeat
the REM AutoFit process.
However, the automatic adjustments from REM AutoFit will still be used to
calculate the prescribed setting if there are changes to anything else, such
as the audiogram, fitting rationale or age.
Print
In Step 4: Finish, you can print the measurements according to your current
on-screen selection. To print after you have closed the REM AutoFit tool, go
to the Genie/Genie 2 top menu and click Print.
Step 4: Finish
Print
Removing measured real ear data