oticon REM AutoFit Quick Manual page 3

Hearing aid fittings

Advertisement

Before using REM AutoFit
Hearing aid settings during measurement
Hearing aid settings during measurement
The following controls are set automatically during REM AutoFit measure-
ments and restored to their original states afterwards, except for Overall
Loudness (Super Power instruments only) which stays on 0.
Noise Management
Feedback Manager
Adaptive Directionality
Adaptation Manager
Overall Loudness
(Super Power
Instruments only)
Speech Rescue
(Super Power
Instruments only)
3
Step-by-step Guide
Step 2: Unaided
Step 1: Calibrate
Genie 16.2
Genie 2 16.2
onwards
onwards
Current setting in
Current setting in
Automatics
OpenSound Navigator
On
On
Off
Off
3
3
0
N/A
Off
N/A
Automatic
Step 3: Aided
Manual
Step-by-step Guide
On the Fitting step in Genie/Genie 2, enter the REM tool from the More
Tools menu in the left taskpane. Launch REM AutoFit from the bottom of the
REM tool.
Step 1: Calibrate
It is necessary to calibrate the probe tube before measurement.
a.
Align the probe tube(s) with the reference microphone(s) and hold 0.5
m from the loudspeaker. Check the REM system's guidance on accept-
able distance.
b.
Click Start on the relevant side to measure one side. To calibrate both
sides, click Both. Depending on the REM system, binaural measurement
may take place simultaneously or sequentially.
c.
The system automatically turns off the signal when the calibration is
complete. If you need to stop the calibration prior to completion, click
Stop.
d.
The result of the tube calibration is displayed on the graphs.
e.
Repeat these steps for each new probe tube.
When you are satisfied with the calibration, proceed to Step 2:
f.
Unaided.