Program type
Single Wavelength
Multi Wavelength
Free programming
6.1.1.1

Single wavelength settings

Table 8 Program descriptions
Description
Readings at a defined wavelength
In the Multi Wavelength mode, absorbance values can be measured at up to four wavelengths
and the results can be mathematically processed to obtain sums, differences and relationships.
A high degree of flexibility results in individual options to create a user program. The free
programming is an expanded form of test or method programming.
The following parameters can be defined when selecting the
single wavelength mode:
Units:
1. Select the required unit from the list.
2. Press Continue.
Note: Units of measure not included in this list can be added in the edit
program under Program Options, Edit. Select Units, Edit and then
New.
Wavelength (single wavelength program type):
1. Enter measured wavelength.
2. Press Continue.
Concentration resolution (number of decimal places)
1. Select the required number of decimal places to be
2. Press Continue.
A wavelength in the range of 190–1100 nm can be entered.
displayed after the decimal point from the displayed list.
Expanded programs
81