Power Management; Lamp Control - Hach DR 6000 User Manual

Laboratory spectrophotometer for water testing
Hide thumbs Also See for DR 6000:
Table of Contents

Advertisement

5.2.8

Power Management

5.2.9

Lamp control

1. Press Power Management in "Instrument Setup".
2. Select Sleep Timer.
3. Select the required time span after which the instrument will
go into energy-saving sleep mode if it is not used (refer to
section 4.4, page
24), and confirm with OK.
4. Select UV Lamp Save.
5. Select the required time span after which the UV lamp will be
powered off if it is not used, and confirm with OK.
Note: To protect the UV lamp, do not switch it on and off frequently.
The UV lamp requires time to warm up.
6. Select Visible Lamp Save.
7. Select the required time span after which the VIS lamp will
be powered off if it is not used, and confirm with OK.
Note: The manufacturer recommends the "Automatic" setting for
this. This means the VIS lamp is only switched on momentarily to
perform the measurement.
With the lamp control the UV lamp and the VIS lamp can be
switched on or off manually.
1. Tap Lamp Control in the "Instrument Setup".
2. Select VIS-Lamp or UV-Lamp.