1.
Pull the top bar upwards with both hands near the left
and right end.
2.
A
To raise the side rail to the middle position (A), stop
pulling upwards after a first audible click.
To raise the side rail to the high position (B), pull
upwards to the upper stop.
3. Push the top bar downwards to ensure it is properly
locked.
If the side rail does not lock properly, perform the
following steps to restore proper function.
Pull up the top bar upwards to its stop, without
pushing the release buttons.
Lower the side rail to its lowest position as
described below.
1567446-H
Lowering the side rail
B
1. Push up and hold the two release buttons.
2. To lower the side rail to the middle position, push the
top bar downwards and let go the two release buttons
after passing the upper locking position.
To lower the side rail to the lowest position, push the
top bar downwards to the lower stop, without releasing
the two release buttons.

3.5 Hand control HB85 / HL85

Hand control HB85 / HL85 is equipped with five buttons to
operate the electrical functions of the bed. The buttons on
hand control HL85 can be locked.
Sitting position
Backrest section
Leg section
Thigh section
Height adjustment
1. Up: press left side of the button (p).
2. Down: press right side of the button (q).
1. Up: press left side of the button (p).
2. Down: press right side of the button (q).
1. Up: press left side of the button (p).
2. Down: press right side of the button (q).
1. Up: press left side of the button (p).
2. Down: press right side of the button (q).
1. Up: press left side of the button (p).
2. Down: press right side of the button (q).