5.7 Basic Program

After the menu has been accessed with ENTER, carry out the
BASIC PROGRAMMING of the unit.

5.8 Program Values

To call up the factory-set basic default setting, hold down the M
and i keys briefly while switching the unit ON.
PROGRAM MENU
Testomat ECO
BASIC PROGRAM
ENTER
PROGRAM VALUES
ENTER
MODE OF OPERAT.
Interval
Type of reagent
Display of unit
Limit values
Flush time
Water meter
BOB operation
Function LV1
Function LV2
Hysteresis LV1
Hysteresis LV2
Function Stop
Interface
Fig. 5.4 Basic Programming Structure
Caution. To reset the factory-set basic default setting, hold down the M and i keys briefly while switching the unit ON.
The previous configuration is erased.
IM/AW101
Rev. C
Values and settings are described in the structure of the basic
program.

5.8.1 Structure of the Basic Programming

Abbreviations: s = seconds m = minutes, h = hours, d = days, l
= liters
TIME CONTROLLED*
Volume interval
TYPE TH2005
TYPE TH2025
*
TYPE TH2100
TYPE TH2250
°
LV1:025 dH
°
LV2:035 dH
FLUSH
10s
FUNCTION ON
Function off
*
DURATION
*
Impulse
Interval
Time
00m:10s
HYSTERESIS LV2 1
TYPE 0-20 mA
Type 4-20 mA
*
5 Menu Structure
TIME
10m
Volume
0001
DISPLAY dH
*
°
Display f °
Display ppm CaC03
Display mmo1/1
1 L/IMPULSE
2.5 l/Impulse
5 l/Impulse
10 l/Impulse
100 l/Impulse
*
500 l/Impulse
1000 l/Impulse
Imp./l
xxx.x
DURATION
*
Impulse
Interval
Two point
Time
00m:10s
HYSTERESIS LV2 1
NORM.OPEN.CONT.
Norm.close con *
13