Alco‐Sensor FST®
 
Alco‐Sensor FST® Battery:  
If the device battery voltage is too low to perform a test sequence, the display will show BAT 
and testing will be disabled, or the device will not turn on. Return the unit to a qualified ASD 
calibrator for battery replacement.  
Service and Repair:  
The Alco‐Sensor FST® requires a yearly service check. For annual service and all repairs (other 
than battery replacement) ship the unit to the authorized service agency: DAVTECH Analytical 
Services Inc.  
 
8.0 Additional Considerations 
8.1 Alternative Reasons for Subject Impairment 
When  alcohol  and  various  drugs  are  taken  in  combination,  unexpected  results  may  occur. 
Combination  effects  are  often  characterized  by  relatively  low  BAC  and  the  presence  of  gross 
symptoms  of  impairment  and  intoxication.  When  alcohol‐drug  interaction  impairment  is 
suspected,  medical  assistance  may  be  required.  Alternatively,  a  Drug  Recognition  Evaluator 
(DRE) may be consulted.  Further investigation may be required. 
In addition, a medical crisis may mimic impairment due to alcohol and/or drugs. 
8.2 Articulation 
Be  prepared  to  articulate  discrepancies  between  the  Alco‐Sensor  FST®  and  evidentiary  test.  
There may be occasions when the Alco‐Sensor FST® result does not correspond with the results 
from  the  subsequent  evidentiary  test.  For  example,  a  FAIL  result  was  obtained  on  the  Alco‐
Sensor FST® but the subsequent evidentiary result was less than 100mg%. These discrepancies 
may be due to:  
1. Time delay between Alco‐Sensor FST® test and evidentiary test 
The  result  obtained  on  the  ASD  may  differ  from  the  result  obtained  on  an  Approved 
Instrument as a result of the time delay between the tests.  As time passes, the breath 
test subject will be eliminating alcohol, resulting in a lower blood alcohol concentration 
at the time of the later breath tests. 
2. Presence of mouth alcohol 
If  mouth  alcohol  is  present,  and  a  15  minute  time  period  is  not  implemented  prior  to 
testing on the ASD, a false high result may be obtained on the ASD, followed by a lower 
result on the approved instrument.  This can be avoided by ensuring that a 15 minute 
wait period occurs when recent consumption of alcohol is suspected. 
 
 
 
 
Alco‐Sensor FST® Operator's Manual 
29 of 30 
Operator's Manual 