4.0 Operate the Alco‐Sensor FST® 
 
4.1 Operational Overview of the ALCO‐SENSOR FST®  
 
It is imperative all ASD operators are able to adequately articulate how the device functions and 
to describe the correct sequence of messages displayed by the device, if called upon to give 
testimony in court. Although some messages are displayed only very briefly, failure to identify 
and articulate these messages can raise concerns about the proper operation of the device.  
 
Commence a test by attaching a clean mouthpiece to the device. Insert the closed end of the 
mouthpiece into the mouthpiece channel, with the flat side of the mouthpiece downward and 
press the mouthpiece into position. The two holes on the underside of the mouthpiece will 
naturally align and attach to the appropriate ports of the Alco‐Sensor FST®. 
  
Press the ON button (labeled with an "I" symbol within a diamond shape) located opposite the 
display for about one second so the display illuminates and an audible beep sounds.  
The backlight of the display is normally green for all functions but turns an amber color when 
displaying a WARN result, or a red color when displaying a FAiL result.  
 
The test sequence starts by briefly indicating the device temperature and battery strength. The 
acceptable operating temperature of the device is ‐12°C to 55°C but the test sequence will 
automatically end if the temperature is not in the acceptable range or if the batteries are too 
weak to complete the test. There is no need to record this temperature.  
 
Following the temperature and battery check, the display will show BlNK and perform a blank 
test to demonstrate the alcohol free status of the device. A blank result other than 0 mg% will 
generate a status message NULL FAiL and end the test sequence automatically, so there is no 
need to record the blank reading. A 0 mg% blank will allow the sequence to continue with WAIT 
appearing on the display while the sequence progresses automatically. 
  
A double beep will sound when the device is ready to accept a sample of breath and there will 
be a small flashing icon of a person's head in the upper left corner of the display and the 
message Blow. The ASD operator should instruct the subject to "take a normal breath in, seal 
your lips around the mouthpiece, and blow continuously through the mouthpiece until I tell 
you to stop". 
  
 
 
Alco‐Sensor FST® Operator's Manual 
12 of 30 
2018‐12‐18 