Alco‐Sensor FST®
 
 
Flow INS indicates that the subject's breath flow rate is inconsistent.  
 
 
Solution:    Instruct  the  subject  to  provide  a  continuous  sample  with  a 
 
moderate rate of breath flow. 
 
 
 
Flow CUT tells the operator that the subject provided enough breath flow to 
capture  a  sample,  but  their  breath  flow  was  stopped  too  abruptly  or  there 
 
has been an attempt to suck back the breath as the sample is being taken for 
 
analysis.  
 
Solution:    Instruct  the  subject  to  provide  a  continuous  sample  with  a 
moderate rate of breath flow. 
 
 
 
 
Flow  HIGH indicates that the subject's breath flow exceeded the maximum 
allowable flow rate. 
 
Solution:    Instruct  the  subject  to  provide  a  continuous  sample  with  a 
 
moderate rate of breath flow. 
 
 
In  each  of  these  examples,  the  subject  is  allowed  three  (3)  attempts  to  provide  an  adequate 
sample before the test is aborted with a Blow Void message on the display and the Alco‐Sensor 
FST® turns itself off. This message will appear after the third failed attempts error message is 
shown.  Instruct  the  subject  to  provide  a  continuous  sample  with  a  moderate  rate  of  breath 
flow. If the subject is to be given the option to attempt another breath sample (after the three 
unsuccessful attempts), you should replace the mouthpiece with a new one. You  will have to 
initiate a new testing sequence by turning on the Alco‐Sensor FST®. Consider retaining the used 
mouthpiece for evidence. 
 
 
 
Refusals:    If  the  officer  forms  the  opinion  that  the  subject  is  refusing  to  provide  a  suitable 
breath sample, retain the mouthpiece for evidence. The officer should make a visual inspection 
of the mouthpiece used by the subject. The officer should also provide a sample into the device 
to demonstrate that the device is capable of accepting a sample. The officer should make note 
that the mouthpieces used were free of obstruction and the device was working properly. 
 
Alco‐Sensor FST® Operator's Manual 
 
 
Blo
 Void – Indicates that the subject has not met the minimum breath 
W
flow requirements after 3 attempts.   
22 of 30 
Operator's Manual 
PROTECTED A 
2018‐12‐18 