Intoximeters Alco-Sensor FST Operator's Manual page 21

Hide thumbs Also See for Alco-Sensor FST:
Table of Contents

Advertisement

Alco‐Sensor FST®
 
4. Breath Sampling: 
 
 
 
 
 
 
 
 
 
 
 
 
In order for the device to accept a sample, the breath test subject must blow with a minimum 
flow  rate,  produce  a  minimum  volume  of  breath,  and  blow  over  a  minimum  period  of  time. 
Once these criteria have been met by the breath test subject, the sample will be accepted and 
analyzed by the device. 
If  the  subject  stops  blowing  before  the  minimum  sampling  requirements  are  met,  then  no 
sample  will  be  taken  for  analysis.  This  will  be  indicated  by  beep  warnings  and  alternating 
messages on the display. In such a case, depending on the circumstances, the subject may be 
offered  another  attempt  to  provide  an  adequate  sample.  When  Blow  comes  back  on  the 
display, you may proceed by telling the subject to blow again. 
The following status messages may be displayed with inadequate samples of breath: 
 
Flow LOW means the subject did not provide a constant breath flow above 
the minimum flow rate required. 
 
The  subject  is  given  three  attempts  to  provide  an  adequate  sample  before 
 
the test is aborted. 
 
Solution:    Instruct  the  subject  to  provide  a  continuous  sample  with  a 
 
moderate rate of breath flow. 
 
Alco‐Sensor FST® Operator's Manual 
When  ready  for  a  sample  of  breath,  the  display  shows  the  icon  of  a 
person's head flashing and Blow. A double beep will be heard when this 
appears  on  the  display.  Instruct  the  subject  to  take  a  normal  breath  in 
and  then  blow  steadily  through  the  mouthpiece  until  you  tell  them  to 
stop. 
"take a normal breath in, seal your lips around the mouthpiece, and blow 
continuously through the mouthpiece until I tell you to stop".  
Once the subject starts to blow with sufficient flow, the icon of the head 
will  stop  flashing  and  a  dash  will  appear  to  the  right  of  the  head. 
Additional  dashes  will  appear  as  the  subject  continues  to  provide  a 
sample. A continuous tone will be heard as the subject is blowing. 
Once three dashes appear a sample will be taken.  It is not necessary for 
the  subject  to  blow  hard,  but  rather  a  steady  and  continuous  sample  is 
best  for  successful  sample  collection.    The  subject  has  up  to  three  (3) 
minutes and/or three (3) attempts in which to provide a suitable sample 
once the Blow message is displayed. 
21 of 30 
Operator's Manual 
PROTECTED A 
2018‐12‐18 