8.2 Steri-Mate
Handpiece
The Steri-Mate handpiece accepts all Cavitron 30K Ultrasonic Inserts. The Cavitron Steri-Mate 360 Handpiece accepts all
Cavitron 30K plastic grip type Ultrasonic Inserts. It is not compatible with the Cavitron 30K Slimline metal grip type inserts
(30K SLI-10S, SLI-10L and SLI-10R).
Handpiece Connector
(Handpiece & Mating
Assembly are keyed)
®
Steri-Mate
360 Handpiece
360 Rotating
Handpiece Nose
Insert Port
360 Rotating Handpiece Nose
To rotate the insert, place fingers on the nose of the handpiece and
rotate to desired position. This allows adjustable hand positioning,
free flowing movement and access within the anterior and posterior
of the oral cavity.
ENGLISH • 8
Cable Assembly
Date Codes
Lavage Flow Adjustment
Holding open end of handpiece upright, lavage can be increased by
holding the handpiece and gently twisting the blue lavage adjustment
knob clockwise. The rate of flow through the Handpiece determines the
temperature of the lavage. Low flow rates produce warmer temperatures,
high flow rates produce cooler temperatures.
BLUE Lavage Knob