DENTSPLY Cavitron SPS Directions For Use Manual

DENTSPLY Cavitron SPS Directions For Use Manual

Ultrasonic scaler with steri-mate handpiece

Advertisement

Introduction
Cavitron
®
SPS
Ultrasonic Scaler
with Steri-Mate
DENTSPLY
Professional is an ISO9001 and EN46000 certified
®
company. The Cavitron
®
SPS
Underwriters Laboratories Inc. with respect to electric shock,
fire, mechanical hazards only in accordance with UL 2601-1 1
and Can/CSA C22.2 NO. 601.1, assigned control # 13VA.
The System operates by converting ordinary AC house current
into high frequency current. The ultrasonic system consists of
two parts: an insert, and the SPS
system incorporates two closed loops. One loop provides
automatic tuning (operating frequency is adjusted to be at
resonance for each insert), the second loop automatically controls
the tip stroke over different working conditions. The Cavitron
SPS
scaler produces 30,000 microscopically small strokes per
second at the insert's working tip. This combined with acoustic
effects of the coolant water, produces a synergistic action that
literally "powers away" the heaviest calculus deposits while
providing exceptional operator and patient comfort.
®
Handpiece
Scaler System is classified by
electronic system. The SPS
Cavitron
Ultrasonic Scaler with
Steri-Mate
Directions For Use
Please read carefully and completely before operating unit.
Technical Support
For technical support and repair assistance in the U.S., call
1-800-989-8826 Monday through Friday, 8:00 AM to 5:00 PM
(Eastern Time). For other areas, contact your local DENTSPLY
representative.
Supplies & Replacement Parts
To order supplies or replacement parts in the U.S., contact your
local Dentsply
®
Distributor or call 1-800-989-8826 or 717-767-8502
Monday through Friday, 8:00 AM to 5:00 PM (Eastern Time). For
®
other areas, contact your local DENTSPLY
Caution: U.S. Federal Law restricts this device to sale by or on
the order of a Dentist.
ENGLISH
SPS
®
Handpiece
®
®
representative.
®
ENGLISH • 1

Advertisement

Table of Contents
loading

Summary of Contents for DENTSPLY Cavitron SPS

  • Page 1 UL 2601-1 1 1-800-989-8826 Monday through Friday, 8:00 AM to 5:00 PM and Can/CSA C22.2 NO. 601.1, assigned control # 13VA. (Eastern Time). For other areas, contact your local DENTSPLY ® representative. The System operates by converting ordinary AC house current into high frequency current.
  • Page 2: Table Of Contents

    Table of Contents Section Section Title Page Number Description of Contents Number Indications ....................................3 Contraindications and Warnings/Adverse Reactions ......................3 2.1 Contraindications 2.2 Warnings Precautions ....................................3 3.1 Precautions for All Systems 3.2 Precautions for Ultrasonic Prophylaxis Procedures Infection Control ..................................4 4.1 Infection Control Information Reference Card 4.2 General Infection Control Recommendations 4.3 Water Supply Recommendations...
  • Page 3: Indications

    Although no instance of pacemaker interference has • For optimum performance use only inserts ever been reported to DENTSPLY , we recommend ® manufactured by DENTSPLY ®...
  • Page 4: Precautions For Ultrasonic Prophylaxis Procedures

    If the installation of your Cavitron System is performed ® ™ Refer to Section 11: Specifications. by someone other than trained DENTSPLY Distributor ® personnel, care should be taken to observe the following 5.4 Unpacking the System requirements and recommendations.
  • Page 5: Power Cord/Power Connection

    Section 5: Installation Instructions cont'd 5.6 Foot Control Cable Assembly 5.5 Power Cord/Power Connection Connection • Verify the Power Control ON/OFF switch located at the left front underside of the System is set to the Align the Foot Control plug with the receptacle on the back of OFF position before proceeding.
  • Page 6: Sps ™ Ultrasonic Scaler Description

    Section 6: SPS Ultrasonic Scaler Description 6.1 System Controls Handpiece Holder Securely holds the System's Handpiece when system is not being used. Top Grooves Handpiece Provide a convenient, temporary Operates all Cavitron ® 30K™ palcement for the handpiece during Ultrasonic inserts and trans- treatment procedures.
  • Page 7: Cavitron Ultrasonic Inserts

    This activates the "Boost" mode. and irrigating water flows. irrigating flow stop. Section 7: Accessories 1. Steri-Mate ® Detachable Sterilizable Handpiece *For detailed information, contact your local DENTSPLY ® Representative or authorized DENTSPLY ® Distributor. 2. AC Power Cord Set 3. Cavitron Ultrasonic Inserts ®...
  • Page 8: Techniques For Use

    • Periodically check the Cavitron ultrasonic insert for wear ® with the DENTSPLY Insert Efficiency Indicator. ® • The use of a saliva ejector or High Volume Evacuator (HVE) is recommended during all procedures.
  • Page 9: Weekly Maintenance

    Ultrasonic Scaler. A 10-pack of replacement filters is available by for 30 seconds. ordering Part Number 90158 from your local DENTSPLY ® distributor. 4. Place a freshly sterilized insert into the handpiece.
  • Page 10: Specifications

    Section 11: Specifications Cavitron ® SPS: 100/115 VAC Systems Footswitch Not for operating theatres Protection Class 1PXO Electrical Voltage 100V Gen 119J Operating Environment 115V Gen 119 Temperature: 15 to 40 Deg. Celsius Current .9/.8 amps @ 100v/115v Relative Humidity: 30% to 75% (non-condensing) Frequency 50/60 Hertz...