7.3 handpiece / Cable
Steri-Mate
®
Handpiece
7.4 Cavitron 30K ultrasonic Inserts
The many styles of Cavitron and Cavitron Bellissima 30K Ultrasonic Inserts are easily interchangeable for various procedures
and applications. See enclosed literature for specific information.
O-Ring
Provides seal for handpiece coolant.
O-ring should be replaced when worn.
Insert Tip
Shape and size of tip
determines access and
adaptation. Preheated Lavage
directed to tip.
Insert Marking
Finger Grip
Manufacturer, Date
(YDDD=Single digit
year and triple digit day
of year), Frequency,
Type, Tip Lot Number
(if applicable)
Lavage Control
Turn the Lavage Control to select flow rate during
system operation. Clockwise increases flow at insert tip,
counterclockwise decreases flow. The flow rate through the
handpiece also determines the temperature of the lavage.
Lower flow rates produce warmer lavage. Higher flow rates
produce cooler lavage.
If the handpiece becomes warm, increase the flow rate.
With experience, the Dental Healthcare Professional will be
able to determine the best flow rate setting for optimum
operating efficiency and patient comfort.
Swivel Feature
Reduces cable drag as handpiece rotates during procedures.
Steri-Mate Grip Accessory (not shown)
The Steri-Mate Grip provides an ergonomic and comfortable
grasp of the handpiece. The grip is sterilizable and is
available in several different colors as an accessory for your
Steri-Mate Handpiece. See installation instructions provided
with the grip.
Connecting Body
Transfers and amplifies mechanical
motion of stack to insert tip.
Magnetostrictive Stack
Converts energy provided by the handpiece
into mechanical oscillations used to activate
the insert tip.
10