DENTSPLY Cavitron G139 Directions For Use Manual

DENTSPLY Cavitron G139 Directions For Use Manual

Built-in ultrasonic scaler with cavitron steri-mate handpiece
Hide thumbs Also See for Cavitron G139:

Advertisement

®
Cavitron
Built-In
G139
Model
Ultrasonic Scaler with
®
®
Cavitron
Steri-Mate
Handpiece
Directions For Use
ENGLISH • 1

Advertisement

Table of Contents
loading

Summary of Contents for DENTSPLY Cavitron G139

  • Page 1 ® Cavitron Built-In G139 Model Ultrasonic Scaler with ® ® Cavitron Steri-Mate Handpiece Directions For Use ENGLISH • 1...
  • Page 2 (operating frequency is adjusted to be at resonance for each insert), the second loop automatically controls the tip stroke over different ® working conditions. The DENTSPLY Cavitron Built-In scaler produces 30,000 microscopically small strokes per second at the insert’s working tip. This, combined with acoustic effects of the coolant water, produces a synergistic action that literally “powers...
  • Page 3: Table Of Contents

    Ultrasonic Scaler System Description ................7–10 8.1 System Controls ® 8.2 Steri-Mate Handpiece ® 8.3 DENTSPLY Cavitron 30K™ Ultrasonic Inserts 8.4 Foot Control Information & Operation Techniques for Use .......................10–12 9.1 Patient Positioning 9.2 Performing Ultrasonic Scaling Procedures 9.3 Patient Comfort Considerations System Care ........................11–12...
  • Page 4: Indications For Use

    Although no instance of interference Disposal of Unit has ever been reported to DENTSPLY, we recommend that the Disposal of unit shall be in accordance with local regulations. handpiece and cables be kept at least 6 to 9 inches (15 to 23 cm) away from any device and their leads during use.
  • Page 5 • the dental procedures being undertaken, accordance with manufacturer’s instructions for a minimum of 5 minutes. • and applicable industry and governmental agency recommendations for infection control in dental • Prior to beginning treatment, patients should rinse with a known antimicrobial such as Chlorhexidine Gluconate 0.12%.
  • Page 6: Precautions

    In general, it is recommended that ultrasonic inserts be discarded and replaced after one year of use to maintain optimal efficiency and avoid breakage. A DENTSPLY Professional Insert Wear Indicator is enclosed for your use. ENGLISH • 5...
  • Page 7: Infection Control

    P.M. (Eastern Time). For areas outside the U.S., contact kPa) maximum. If your dental water system's supply line your local DENTSPLY Professional representative. pressure is above 60 psi, install a water pressure regulator on the water supply line to your Ultrasonic Scaler.
  • Page 8: Ultrasonic Scaler System Description

    Water flow through the handpiece is adjusted by rotating a control Description element on the cable connector (blue). 8.1 System Controls The DENTSPLY Cavitron® Built-In module is enabled whenever the clinician removes the handpiece from the handpiece holder. ON/OFF Function The Ultrasonics is switched ON and OFF using the treatment unit’s...
  • Page 9: Steri-Mate ® Handpiece

    ® 8.2 Steri-Mate Handpiece The Steri-Mate handpiece accepts all Cavitron 30K Ultrasonic Inserts. The Cavitron Steri-Mate 360 Handpiece accepts all Cavitron 30K plastic grip type Ultrasonic Inserts. It is not compatible with the Cavitron 30K Slimline metal grip type inserts (30K SLI-10S, SLI-10L and SLI-10R).
  • Page 10: Dentsply Cavitron ® 30K™ Ultrasonic Inserts

    ® 8.3 DENTSPLY Cavitron 30K™ Ultrasonic Inserts ® The many styles of DENTSPLY Cavitron Ultrasonic Inserts are easily interchangeable for various procedures and applications. O-Ring Connecting Body Provides seal for handpiece coolant Transfers and amplifies mechanical motion of the stack to insert tip.
  • Page 11: Foot Control Information & Operation

    Vacuum Evacuator (HVE). 9.2 Performing Ultrasonic Scaling Procedures ® • The edges of DENTSPLY Cavitron Ultrasonic Inserts are intentionally rounded so there is little danger of tissue laceration with proper ultrasonic scaling techniques. The lips, cheek and tongue should be retracted to prevent accidental prolonged contact with the activated tip whenever the insert is placed in the mouth.
  • Page 12: Patient Comfort Considerations

    Clean and disinfect the handpiece cable assembly by 1. Switch ON the dental treatment unit (refer to instruction applying a medically approved non-immersion type manual of treatment unit). disinfectant solution* carefully following the instructions ® 2. Remove the DENTSPLY Cavitron cable connector ENGLISH • 11...
  • Page 13 Allow the disinfectant solution to air dry. Shut down at the end of the day Flush the DENTSPLY Cavitron Built-In ultrasonic system Clean and disinfect all surfaces of the dental treatment according the dental treatment unit manufacturer’s control head according to the manufacturer’s...
  • Page 14: Troubleshooting

    Section 11: Troubleshooting Some basic troubleshooting procedures that will help avoid unnecessary service calls are listed below. wear indicator ENGLISH • 13...
  • Page 15 ENGLISH • 14...
  • Page 16 ENGLISH • 15...
  • Page 17 Verify input water temperature is not higher than F (25 ENGLISH • 16...
  • Page 18: Warranty

    Section 12: Warranty Period Output Frequency 30 kHz Power 3 – 30 Watts ® The Cavitron Built-In G139 Ultrasonic Scaler is warranted Water Flow Rate 10 > flow > 60 ml/min ® for TWO YEARS from date of purchase. The Steri-Mate Storage and Shipping Conditions: Handpiece enclosed with your system is warranted for SIX Ambient temperature range:...
  • Page 19: Symbol Identification

    Section 15: Symbol Identification Type B Equipment Consult Instructions AC power (AC 24~) For Use MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1CAN/CSA-C22.2 No. 601.1, ANSI/AAMI ES60601-1 (2005, 3rd ed.) CAN/CSA-C22.2 No. 60601-1 (2008), 13VA This symbol is a mandatory marking for devices enter- ing the European market to indicate conformity with the...
  • Page 20 DENTSPLY DeTrey Italia S.r.l. 161 Vinyl Court Bangkok 10120 Via Curtatone, 3 Woodbridge, Ontario 00185 Rome, Italy Canada L4L 4A3 DENTSPLY Mexico, S.A. de C.V. Calzada Vallejo No. 846 Col Industrial Vallejo 02300 Mexico, D.F. Printed in U.S.A. Form #88016 Rev. 1 (07/15) Manufactured by: Imported and DENTSPLY DeTrey GmbH DENTSPLY Professional Distributed by: De-Trey-Str. 1 1301 Smile Way DENTSPLY Canada 78467 Konstanz York, PA 17404-1785 Woodbridge, Ontario Germany L4L 4A3...

Table of Contents