Advertisement

seca 869 UK
GB
Instruction manual and guarantee
2

Advertisement

Table of Contents
loading

Summary of Contents for Seca 869

  • Page 1 869 UK Instruction manual and guarantee...
  • Page 2 English By purchasing the seca 869 electronic In addition to the conventional determina- tion of weight, the seca 869 also has a personal scale, you have acquired a high- ly-accurate and sturdy piece of equip- function for determining body mass index.
  • Page 3: Desk-Top Stand

    • 6 mignon-type AA 1.5 V batteries Desk-top stand If you prefer, you can stand the display part rather than hang it on the wall: Alter the angle of the display part to enable the display to be read most easily. Model 869...
  • Page 4: Fitting The Wall Bracket

    Fitting the wall bracket The display part is intended for mounting on the wall. Use the mounting material supplied to do so. – Screw the wall bracket to the wall as shown. Be aware that you have a length of 2 m of cable between the scale and the dis- play part.
  • Page 5: Power Supply

    – Then turn the scale back up the right way. Note: the scale is prepared for use with a seca mains unit which is availa- Connection for mains unit ble as an accessory. Connecting the connecting cable to the weighing platform...
  • Page 6: Setting Up And Aligning The Scale

    Setting up and aligning the scale The screwed feet can be used to level off any slight unevenness in the floor. – Place the scale on a level surface. spirit level – Align the scale by screwing the foot screws. The air bubble in the spirit level must be right in the centre of the circle.
  • Page 7: Operation

    1.5 kg. The display flashes until a stable weight has been measured. The display is then frozen until the next weighing op- eration. HOLD appears in the display. – Read the weighing result off the digital display. Model 869...
  • Page 8: Switching The Weight Display

    – Ask the patient to step off the scale. 0. 0 0 The weighing result has been erased. The scale is ready for the next patient. – If a load greater than 250 kg, 550 lbs or St0P 39 sts is placed on the scale, pears in the display.
  • Page 9: Store Weight Value (Hold)

    8. 8 . 8 . 8 . 8 . 0. 0 appear con- secutively in the display. The scale is then automatically set to zero and ready for operation. – Press the bmi key. On the display, you will see the last height entered. Model 869...
  • Page 10: Evaluating Body Mass Index For Adults

    – You can use the arrow keys to adjust the value in increments according to the graduations displayed. – Once the correct value is set, press the bmi key again. The BMI function is now activated. – Step on the scale and keep still. –...
  • Page 11 Service department. 7. Servicing Your seca scale leaves the factory with an accuracy greater than ± 0.15 %. To ensure that this accuracy is maintained, the product must be carefully set up and regularly maintained.
  • Page 12: Changing The Batteries

    Changing the batteries bAtt When appear in the display, you should change the batteries. You need 6 type AA 1.5 V batteries. – Carefully turn over the scale so that the base of the scale is accessible. – Open the cover. –...
  • Page 13: Spare Parts And Accessories

    Switchmode power adapter, 100-240V~ / 50-60Hz / 12V= / 0.5A seca no. 68-32-10-265 Attention! Use only seca original plug-in mains units with 9 V or controlled 12 V output voltage. Commercially-available mains units may supply a higher voltage than stated and thus lead to damage to the scale.
  • Page 14: Warranty

    12. Declaration of conformity seca gmbh & co. kg hereby declares that the product meets the terms of the applicable European directives. The unabridged declaration of conformity can be found at: www.seca.com.

Table of Contents