  • Page 1 869 UK Instruction manual and guarantee...
  • Page 2 English By purchasing the seca 869 electronic In addition to the conventional determina- tion of weight, the seca 869 also has a personal scale, you have acquired a high- ly-accurate and sturdy piece of equip- function for determining body mass index.
  • Page 3: Desk-Top Stand

    3. Before you start… Remove the packaging and place the scale on a firm, level surface. The scope of supply includes: • Weighing platform • Display head plus connecting cable • Wall bracket • 2 Screws • 2 Rawlplugs • 6 mignon-type AA 1.5 V batteries Desk-top stand If you prefer, you can stand the display part rather than hang it on the wall:...
  • Page 4: Fitting The Wall Bracket

    Fitting the wall bracket The display part is intended for mounting on the wall. Use the mounting material supplied to do so. – Screw the wall bracket to the wall as shown. Be aware that you have a length of 2 m of cable between the scale and the dis- play part.
  • Page 5: Power Supply

    – Then turn the scale back up the right way. Note: the scale is prepared for use with a seca mains unit which is availa- Connection for mains unit ble as an accessory. Connecting the connecting cable to the weighing platform...
  • Page 6: Setting Up And Aligning The Scale

    Setting up and aligning the scale The screwed feet can be used to level off any slight unevenness in the floor. – Place the scale on a level surface. spirit level – Align the scale by screwing the foot screws. The air bubble in the spirit level must be right in the centre of the circle.
  • Page 7: Operation

    4. Operation Controls and displays The controls and display are in the display part. Switch scale on and off  switch between kg and lbs press briefly  switch between lbs and sts hold down in function menu  increase value ...
  • Page 8: Switching The Weight Display

    – Ask the patient to step off the scale. 0. 0 0 The weighing result has been erased. The scale is ready for the next patient. – If a load greater than 250 kg, 550 lbs or St0P 39 sts is placed on the scale, pears in the display.
  • Page 9: Store Weight Value (Hold)

    0. 0 The display is reset to zero, flashes in the display until weight is detected to be stable. Only then is the Tare function fully activated. The note “NET” appears. – Now take the additional load onto the scale. You can leave the scale to do so, ---- in which case appears.
  • Page 10: Evaluating Body Mass Index For Adults

    – You can use the arrow keys to adjust the value in increments according to the graduations displayed. – Once the correct value is set, press the bmi key again. The BMI function is now activated. – Step on the scale and keep still. –...
  • Page 11 Service department. 7. Servicing Your seca scale leaves the factory with an accuracy greater than ± 0.15 %. To ensure that this accuracy is maintained, the product must be carefully set up and regularly maintained.
  • Page 12: Changing The Batteries

    Changing the batteries bAtt When appear in the display, you should change the batteries. You need 6 type AA 1.5 V batteries. – Carefully turn over the scale so that the base of the scale is accessible. – Open the cover. –...
  • Page 13: Spare Parts And Accessories

    Switchmode power adapter, 100-240V~ / 50-60Hz / 12V= / 0.5A seca no. 68-32-10-265 Attention! Use only seca original plug-in mains units with 9 V or controlled 12 V output voltage. Commercially-available mains units may supply a higher voltage than stated and thus lead to damage to the scale.