WARNINGS
Indicates a potentially hazardous situation, if not avoided, could result in death or serious injury.
» Before use, all individuals who will be using this product must become thoroughly
familiar with the information contained in this Operation Manual. Strict adherence to
the operating instructions is necessary for safe and effective product performance.
This product will perform only as designed and only if installed and operated in
accordance with the manufacturer's operating instructions.
» Although the sensor of this device has been tested with various anesthesia gases including
nitrous oxide, Halothane, Iso urane, En urane, Sevo urane and Des urane and found to
have acceptably low interference, the device in entirety (including electronics) is not suitable
for use in the presence of a ammable anesthetic mixture with air or with oxygen or nitrous
oxide. Only the threaded sensor face, ow diverter, and T adapter may be allowed to
contact such a gas mixture.
» This device does not contain automatic barometric pressure compensation.
» Do not use near any type of flame or flammable/explosive substances, vapors or
atmosphere.
» Not for use in an MRI environment.
» Improper use of this device can cause inaccurate oxygen readings which can lead to improper
treatment, hypoxia or hyperoxia. Follow the procedures outlined in this user manual.
CAUTION:
Indicates a potentially hazardous situation, if not avoided, could result in minor or moderate
injury and property damage.
» The Maxtec MAX-250 oxygen sensor is a sealed device containing a weak acid electrolyte,
lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and
should be disposed of properly, or returned to Maxtec for proper disposal or recovery.
» The Handi
+ is not intended for steam, ethylene oxide or radiation sterilization. Do not
®
autoclave or expose the sensor to high temperatures. Do not immerse the Handi
analyzer in any cleaning solution.
» The ow diverter provided with the Handi
the diverter when performing static sampling (e.g., in incubators, oxygen tents, oxygen
hoods).
» Do not attempt any repairs or procedures, which are not described in this Operation Manual.
Maxtec cannot warrant this product from damage resulting from misuse, unauthorized repair
or improper maintenance of this product.
» The Federal (USA) law restricts this device to sale by or on the order of a physician.
» There are no internal user-serviceable parts.
» Never install the sensor in a location that will expose the sensor to patient's exhaled breath
or secretions.
866.4.Maxtec
II
+ is for use with owing gases only. Do not use
®
www.maxtecinc.com
+ oxygen
®
Need help?
Do you have a question about the Handi+ and is the answer not in the manual?