Model 2701 Ethernet-Based DMM / Data Acquisition System Service Manual
Assembly drawings
Use the following assembly drawings to assist you as you disassemble and reassemble the 2701.
Also, refer to these drawings for information about the Keithley Instruments part numbers of most
mechanical parts in the unit.
•
Front Panel Assembly
•
Power Module / Card Cage Assembly
•
Power Module / Transformer / Chassis Assembly
•
Front Panel / Card Cage / Chassis Assembly
•
Chassis Assembly
•
Final Chassis Assembly
•
Final Inspection
Disassembly procedures
Case cover removal
Before removing the case cover, disconnect the line cord and any test leads from the
instrument.
To remove the handle, mounting ears, rear bezel, bottom screws, and cover:
1. Swing the handle below the bottom surface of the case and back until the orientation arrows on
the handles line up with the orientation arrows on the mounting ears. Pull the ends of the handle
away from the case.
2. Remove the screw that secures each mounting ear, then pull down and out on each mounting ear
to remove.
When re-installing the mounting ears, be sure to mount the right ear to the right side of the chassis
and the left ear to the left side of the chassis. Each ear is marked RIGHT or LEFT on its inside
surface.
3. Loosen the two captive screws that secure the rear bezel to the chassis, then pull the bezel away
from the case.
4. Remove the four bottom screws that secure the case to the chassis.
5. Grasp the front bezel of the instrument and slide the chassis forward.
6. Slide the chassis out of the metal case.
2701-902-01 Rev. B / December 2018
Section 5: Disassembly
5-3