Measurement - Duration - Malvern Mastersizer 3000 User Manual

Table of Contents

Advertisement

Making measurements

Measurement - Duration

The Measurement Duration settings can be used to separately specify both the red
and blue light measurement durations.
Mastersizer 3000
Section/option
Description
Red measure-
Background measurement duration (seconds)
ment duration(s)
This facility will make a measurement using only clean dis-
persant – this is then subtracted from the sample measure-
ment by the system in order to 'clean' the data. This is often
set to the same duration as the sample measurement.
Sample measurement duration (seconds)
The optimum measurement time depends on the size of the
sample, its particle size distribution and the dispersion
accessory used. If a material is monomodal its essential par-
ticle size characteristics can be captured in less time than a
material with a broad particle size distribution, which will
need to be measured longer to ensure that the coarser parti-
cles have been properly represented.