Replacements; Skin Replacement - Gaumard S402.100 User Manual

Smash advanced patient training arm
Table of Contents

Advertisement

Replacements

SKIN REPLACEMENT

To replace the skin of training arm follow the
instructions below:
1. Drain any fluids from prior training exercises;
refer to the "Draining The System" section to
empty the system.
2. Close all clamps.
3. Disconnect the tubes from all three ports
located on the arm.
4. Remove the dispensing bag from the metal
stand.
5. Disconnect wrist band with the snaps
holding the training arm in place.
6. Use the Allen key provided to remove the
upper bolt from the shoulder.
7. Gently lift the arm and release it completely
from the base.
18 |
User Guide | Appendix |
8. Keep the shoulder base pad and the nut in
position as shown below.
9. Remove the arterial and venous or fistula
insert
10. Remove the incision and suture insert.
11. Gently remove skin starting from the
shoulder.