ABB Aztec 600 User Manual page 45

Colorimetric and ion-selective analyzers
Hide thumbs Also See for Aztec 600:
Table of Contents

Advertisement

Aztec 600
Colorimetric and ion-selective analyzers
Octet
Bit
Description
DIAGNOSIS
1
0
Hardware failure of the electronics*
1
Not used
2
Not used
3
Electronics temperature too high
4
Memory error
5
Failure in measurement
6
Not used
7
Not used
2
0
Not used
1
Power supply failed
2
Configuration not valid
3
Not used
4
Not used
5
Maintenance required**
6
Not used
7
Not used
3
Not used
4
Not used
Table B.2 Diagnostic Data (Diagnosis)
*Hardware failure may be one of the following:
1Head Heating Failure
2 Head Temperature Critical (> 60 °C)
3 Head LED Calibration Failure
4 Head LED Always Dark
5 Head LED Always Light
6 Head Motor/Piston Fault
7 Head LED Origin Fault
8 Head Communications Failure
9 PSU Supply Current Too High
10 Internal Communications Failure
**Maintenance required may be one of the following:
1 Sample Flow Failure – Colorimetric Stream 1
2 Sample Flow Failure – Colorimetric Stream 2
3 Sample Flow Failure – Colorimetric Stream 3
4 Sample Flow Failure – ISE
IM/AZT6PBS–EN Rev. D
Appendix B – Acyclic Parameter Mapping
Diagnosis Mask ( 1 = Enabled)
1
0
0
1
1
1
0
0
0
1
1
0
0
1
0
0
0
0