In order to be able to use certain functions of this instrument (in particular the measurement data management), you need to accept the testo Cloud terms and conditions of use, which you will find under the login of the respective testo solution at www.testo.com/login.
2 Safety and waste disposal 2.1 Intended purpose The testo Saveris Food Logistics Inbound data logger is used to store and read out series of measurements during the transportation of products that are subject to cold chain requirements. .
At the end of its useful life, dispose of the instrument via separate collection for electrical and electronic devices. Please observe local regulations regarding waste disposal, or alternatively return the product to Testo for disposal. WEEE Reg. No. DE 75334352 3 Description of the instrument 3.1 Scope of delivery...
4 Product use 3.3 Operating elements of the product Element Element 1 Serial number/barcode 2 Start/Check button (removable) 3 LED (red) 4 LED (green) 5 Serial number/barcode 6 Order number and configuration (integrated) 7 Refer to instruction manual 4 Product use 4.1 Commissioning Switch the logger on.
Page 9
4 Product use The logger’s green and red LEDs flash briefly to acknowledge successful activation. LED flashes green: system test successful Data logger starts recording reading LED flashes red: system test unsuccessful Data logger must not be used. Each time the Start/Check button is pressed again, this produces a...
Page 10
4 Product use Enclose data logger with the shipment. Optional: attach location tape to the outside of the packaging. Optional: attach location sticker to the packaging so that it is clearly visible from the outside.
4 Product use 4.2 Reading out data Remove the data logger and immediately press the Start/Check button. While the button is pressed, the data logger indicates whether the limit values were adhered to/violated. - LED lights up green: limit values adhered to - LED lights up red: limit values violated CAUTION...
The used data loggers can be assembled in the box provided and returned to Testo. Please clean the loggers before assembling them in the box. 4.3 LED status display The following statuses are indicated via the LEDs:...
The data logger does not have an integrated temperature display. Once the measurement data is transferred, it can be viewed in the database. For more detailed information, please consult the documentation for the testo Saveris Food Logistics Inbound software. 4.5 Alarms The LEDs on the data logger provide an indication as to whether the limit values were adhered to or violated.
5 Technical data 5 Technical data Feature Value Measuring range -30 °C to +50 °C / -22 °F to +122 °F Accuracy ± 0.5°C/°F Resolution 0.1 °C/°F Sensor type Digital temperature sensor (semi-conductor basis) Measurement channels 1 (internal) Measurement parameters Temperature (°C, °F) Storage temperature -40 °C to +85 °C /-40 °F to +185 °F...
6 Authorizations 6 Authorizations The use of the wireless module is subject to the regulations and stipulations of the respective country of use, and the module may only be used in each case in countries for which a country certification has been granted.
Page 16
6 Authorizations Europa + EFTA The EU Declaration of Conformity can be found on the testo homepage www.testo.com under the product specific downloads. EU countries: Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE), Estonia (EE), Finland (FI), France (FR), Greece (GR), Ireland...
Page 17
6 Authorizations IC Warning RSS-Gen & RSS-247 statement: This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Page 18
6 Authorizations FCC warning statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.