Maintenance

The Insight BIO does not require regular maintenance. Change the eyepiece lenses only
in a reasonably dust-free environment. If you should require service on your product,
please contact Welch Allyn.

Clean and disinfect the instrument

1. Remove the rechargeable battery from the Insight BIO headband.
2. Clean the Insight BIO's lens glass with a soft, lint-free cloth lightly dampened with
70% isopropyl alcohol. Do not excessively wet the lens glass.
3. Clean and disinfect the remaining surfaces of the Insight BIO with a soft cloth and
70% isopropyl alcohol or germicidal wipes labeled for healthcare equipment. Follow
wipe manufacturer's instructions for appropriate use and contact time, and heed all
warnings and cautions. Do not excessively wet the instrument or allow liquid to
penetrate the optical unit.
4. Clean dirty on/off magnet contacts with cotton swab soaked in alcohol.

Charge the battery

You can charge the battery when it is disconnected from the Insight BIO and also when
it is mounted on the Insight BIO headband.
CAUTION Do not autoclave, immerse, excessively wet, or use germicidal
cleaners other than those identified. Doing so will damage the instrument.
CAUTION The battery charger for the Insight BIO is for charging the #
75360 battery. Never use it for other purposes or on other manufacturers'
batteries.