E8257DS02-S15

Before Applying Power

Verify that the premises electrical supply is within the range of the instrument. The instrument has an
autoranging power supply.
WARNING
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
User's Guide E8251-90283
This Manual:http://www.manuallib.com/agilent/e8257ds02-s15-source-modules-user-guide.html
If this product is not used as specified, the protection provided by the equipment could be impaired.
This product must be used in a normal condition (in which all means for protection are intact) only.
The Mains wiring and connectors shall be compatible with the connector used in the premise
electrical system. Failure, to ensure adequate earth grounding by not using the correct components
may cause product damage, and serious injury.
Always use the three prong AC power cord supplied with this product. Failure to ensure adequate
earth grounding by not using this cord may cause product damage and the risk of electrical shock.
This product is designed for use in Installation Category II and Pollution Degree.
Before switching on this instrument, make sure the supply voltage is in the specified range.
Verify that the premise electrical voltage supply is within the range specified on the instrument.
Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of
the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less
than the maximum operating temperature of the instrument by 4 °C for every 100 watts dissipated in
the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection
must be used.
Safety and Regulatory Information
7