Before You Start - Agilent Technologies CrossLab RapidFire 400 Installation Checklist

System temperature control unit
Hide thumbs Also See for CrossLab RapidFire 400:

Advertisement

Before you start

Ensure the power is turned off and the elevator stacks are empty.
Remove the rear and side panels using a 4-mm hex wrench to loosen the captive screws
holding the panels to the frame.
Remove the right-front door by compressing the hinge fingers and rotating them into the
locked position.
Removing the cables, drawer, and elevator base
1
Remove the door handle M4 x 18-mm screws on the drawer with a T-20 driver and set aside
to reinstall later.
2
Disconnect the three ribbon cables from the back of the robot control box for the front,
middle, and rear elevator stacks.
3
Remove the ferrites on the cables with the ferrite key and set aside for reinstallation later.
4
Using a T-10 driver, remove the three M3 screws on the rear cover over the arm bracket and
remove the M3 screw holding the cable arm bracket to the elevator mounting plate. Set the
cover and fasteners aside for reinstallation later.
5
Feed the ribbon cables up through the access hole in the deck and tie the ribbon cables and
the cable arm to the rear elevator stack, using the long cable ties included in the installation
kit, such that the cable arm bracket is tucked out of the way when the drawer is opened.
6
Using a T-20 driver, remove the drawer handle. Set the handle and fastener aside for
reinstallation later.
7
Clear a space to set the drawer while installing the temperature enclosure and remove the
drawer:
a
Standing in front of the instrument, open the drawer fully.
b
Locate the slide member release tabs in the middle of the slide.
c
With the drawer at its extended position, push the right-slide tab upwards and pull the left-
slide tab downwards to release the drawer.
d
Pull the drawer fully out, supporting the weight as it falls from the end of the slides.
8
Using a T-20 driver, remove the four M4 screws holding the elevator mounting plate to the
base plate.
9
Remove the elevator mounting plate and the O-ring cups and O-rings that were under the
attachment screws and set them aside to be used when reinstalling the elevator mounting
plate.
Revision: A.00, Issued:
January
17,
Document part number:
D0007550
© Agilent Technologies, Inc.
2021
2021
Page ____ of ____
Instrument Installation Checklist