Operation Verification
Checking the Output Power
Table 3-6
Recommended Equipment for the E8663D
E8663D
E8663D Option
Option
509
503
X
X
X
X
X
X
X
X
Perform the following procedure to check the signal generator output power:
1. Zero and calibrate the power sensor to the power meter as shown:
2. Connect the equipment as shown:
3. Press Preset.
4. Press RF On/Off. The RF On annunciator is now displayed.
5. Press Mod On/Off. The Mod Off annunciator is now displayed.
32
Test Equipment
Power Sensor,
Input: APC 3.5 (m)
Power Sensor,
Input: 2.4 mm (m)
Adapter,
3.5 (f) to (f)
Adapter,
3.5 (f) to Type-N (m)
Keysight E8257D/67D & E8663D PSG Signal Generators Installation Guide
Recommended Keysight Test Equipment
Keysight E4413A E-Series power sensor
Keysight 8487A Option H84 power sensor (Option
H84 adds a Standards Lab Calibration)
Keysight 1250-1749 APC 3.5 (f) to
(f) adapter
Keysight 1250-1744
APC 3.5 (f) to Type-N (m) adapter