7-Step Quick Reference Guide
3M
Attest
390 Auto-reader and 3M
TM
TM
ATTENTION: Wear safety glasses and gloves. DO NOT place another label or indicator tape on the vial or on the cap of the BI.
1
2
3
4
5
6
7
3M
Attest
390
TM
TM
Ŕ
Auto-reader
Refer to 3M
Attest
Rapid Readout Biological Indicator (BI) package inserts for complete instructions.
TM
TM
Refer to 3M
Attest
390 Auto-reader Operator's Manual for complete instructions.
TM
TM
Available in Canada from:
3M Infection Prevention Division
3M Canada Company
P.O. Box 5757
London, Ontario N6A 4T1 Canada
1 800 563-2921
File Name : 1708-09713 _Revise390_6StepGuide.indd
File Description : Attest Quick Step 1 sheet (Cncpt #2 Front)
File Size : 11" W x 17" H plus bleed and trim
Colour : 4 Colour Process
Attest
Rapid Readout Biological Indicator (BI)
TM
TM
Ŕ
Allow BI to cool at least 10 minutes.

Cool

Ŕ
Crushing or excessive handling of the BI before cooling may cause
the glass ampoule to burst.

Close

Ŕ
While holding BI in hand, press the cap down.

Crush

Ŕ
Hold BI by the cap and use "Crusher Wall" to crush the BI only one time.
Ŕ
DO NOT use your fingers to crush the glass ampoule.
Ŕ
Hold the BI by the cap and tap bottom of vial on countertop until

Tap

the media wets spore strip at bottom of vial.
Ŕ
DO NOT roll BI between fingers.
Ŕ
Place BI into incubation well that corresponds with colour-coded
BI cap. Observe LCD panel to confirm BI is correctly seated.

Incubate

Remaining incubation time, in minutes, will be displayed directly
below BI on the LCD display to indicate incubation is in progress.
DO NOT REMOVE BI from the incubation well until the
Ŕ
(+) or (-) symbol on the LCD panel indicates the test is complete.
NOTE: Do not reincubate the same BI for a second reading as any subsequent readings would be considered invalid.
Interpret
Ŕ
Record result for both test BI (processed) and positive control (unprocessed).
Results

Record

Ŕ
Record result for both test BI (processed) and positive control (unprocessed).
3M Center, Building 275-4W-02
St. Paul, MN 55144-1000 USA
1 800 228-3957
3M.com/HealthCare
3M
Attest
Rapid
TM
TM
Readout Biological
Ŕ
Indicator (BI)
Unprocessed
Controlled
Processed
Incubated
Incubated
3M, 3M Science. Applied to Life. and Attest
are trademarks of 3M. Used under license in
Canada. Please recycle. Printed in Canada.
© 2017, 3M. All rights reserved. 1708-09713-E
Processed

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Attest 390 and is the answer not in the manual?

Questions and answers

Summary of Contents for 3M Attest 390

  • Page 1 390 Auto-reader Operator’s Manual for complete instructions. Available in Canada from: 3M Center, Building 275-4W-02 3M, 3M Science. Applied to Life. and Attest 3M Infection Prevention Division St. Paul, MN 55144-1000 USA are trademarks of 3M. Used under license in...
  • Page 2 390. Produits offerts au Canada par la : 3M Center, Building 275-4W-02 3M, 3M Science. Au service de la Vie., et Attest sont des Division des solutions pour la St. Paul, MN 55144-1000 USA marques de commerce de 3M, utilisées sous licence au prévention des infections de 3M...

Table of Contents