Guidance and manufacturer's declaration
Guidance and manufacturer's declaration - Electromagnetic immunity
The device KINETEC
SPECTRA Essential is intended for use in the electromagnetic environment specified below. The customer or the user
®
of the device KINETEC
Immunity test
Conducted RF interference
IEC 61000-4-6
Radiated RF interference
IEC 61000-4-3
NOTE 1
At 80 and 800 MHz, the highest frequency range is applicable.
NOTE 2
These directives cannot be applied in every situation. Electromagnetic propagation is affected by absorption and reflection
by structures, objects and people.
The field intensity of fixed emitters such as base stations for radio-telephones (cellular/cordless) and land mobile radios, amateur
a
radio, AM/FM radio broadcasts and TV broadcasts cannot be predicted exactly in theory. To evaluate the electromagnetic
environment due to fixed RF emitters, an on-site electromagnetic investigation should be considered. If the field intensity measured
where the device KINETEC
device KINETEC
SPECTRA Essential machine should be monitored to check that it is working normally. If abnormal results are
®
observed, additional measures may be necessary, such as reorienting or repositioning the device KINETEC
Over the frequency range 150 kHz to 80MHz, field intensities should be less than 3V/m.
b
Recommended separation distances between mobile and portable RF communication devices and the device KINETEC
Essential machine
The device KINETEC
SPECTRA Essential machine is designed to be used in an electromagnetic environment in which radiated RF
®
interference is controlled. The customer or user of the device KINETEC
interference by maintaining a minimum distance between mobile and portable RF communication devices (emitters) and the device
KINETEC
SPECTRA Essential machine, as recommended below, according to the communication device's maximum output power.
®
Maximum assigned output power
for the emitter
W
0.01
0.1
1
10
100
For emitters whose assigned maximum emitted power is not given above, the recommended separation distance d in metres (m) can
be estimated using the equation applicable to the emitter frequency, where P is the emitter's maximum emission power characteristic in
watts (W), according to the latter's manufacturer.
NOTE 1
At 80 and 800 MHz, the separation distance for the highest frequency range is applicable.
NOTE 2
These directives cannot be applied in every situation. Electromagnetic propagation is affected by absorption and
reflection by structures, objects and people.
Spectra Essential
SPECTRA Essential should ensure that it is used in such an environment.
®
Test level according to
EIC 60601
3 Veff
from 150 kHz to 80 MHz
3 V/m
from 80 MHz to 2.5 GHz
SPECTRA Essential machine is used exceeds the aforementioned applicable RF compliance level, the
®
from 150 kHz to 80 MHz
d = 1.2
0.12
0.38
Compliance
level
Mobile and portable RF communication devices should not
be used closer to any part of the device KINETEC
Essential machine, including its cables, than the
recommended separation distance, calculated based on the
equation applicable to the emitter's frequency.
Recommended separation distance
P
3 V
d = 1.2
P
d = 1.2
3 V/m
P
d = 2.3
where P is the emitter's maximum output power characteristic
in watts (W), according to the emitter's manufacturer, and d is
the recommended separation distance in metres (m).
The field intensities of fixed RF emitters, determined by an on-
site electromagnetic investigation
compliance level in each frequency range
There may be interference near appliances bearing the
following symbol:
SPECTRA Essential machine can help prevent electromagnetic
®
Separation distance according to the emitter's frequency
80 MHz to 800 MHz
P
d = 1.2
1.2
3.8
12
EN
Electromagnetic environment –
guidance
from 80 MHz to 800 MHz
from 800 MHz to 2.5 GHz
, should be below the
a
SPECTRA Essential.
®
m
from 800 MHz to 2.5 GHz
P
d = 2.3
0.12
0.38
1.2
3.8
12
SPECTRA
®
.
b
SPECTRA
®
P
0.23
0.73
2.3
7.3
23
12/12

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents