QUANTiC: Readhead dimensions
Reference mark selector sensor position
R>30 Dynamic bend radius
R>10 Static bend radius
*
Extent of mounting faces
QUANTiC linear installation guide
18
29
7.8
P and Q limit switch sensor position
6 min
11.5
Optical centreline marker
Optical centreline
(Incremental and reference mark)
7.8
Mounting faces
Set-up LED
35
23
2 off mounting holes M2.5 through,
counterbored Ø3 × 2.3 deep both sides.
NOTE: The recommended thread engagement is 5 min
(7.5 including counterbore) and the recommended
tightening torque is between 0.25 and 0.4 Nm.
2.5
Dimensions and tolerances in mm
*
8.75
13.5
4.25
4.15
10
15